ARCHTECH™ IOS PROTOCOL

5 Steps for Reliable Implant Scans

5 step scanning protocol

This technique is validated to produce precise and reliable intraoral scans

Scanner tip sweeping the occlusal surface of an arch fitted with ARCHTECH scan posts
01

Occlusal sweep

Starting from the most distal area scan the occlusal aspect in a single, slow, uninterrupted pass.

Scanner held at an angle along the buccal side of the scan posts
02

Buccal sweep

Hold the scanner ~45 ° to the occlusal plane and scan buccally from distal to distal.

Scanner looping around the lingual side of the arch
03

Lingual sweep

Loop around to the lingual, ensuring full sulcus depth and the entire palate is captured.

Close-up of the scanner detailing two scan posts joined by the ScanChain
04

Fill gaps and detail the scan posts

Check the digital model for gaps/missing data; rescan any voids or incomplete areas.

Scanner over the opposing arch
05

Opposing arch

Repeat steps 1-3 for the opposing arch.

Helpful tips

Inspect scanner

Always inspect and (if necessary) calibrate your scanner prior to use. Aged, damaged or dirty tips will reduce their accuracy.

Follow software workflow

Use the dedicated implant workflow in the software.

Use ambient lighting

Optimise ambient lighting for intraoral scanning and avoid strong illumination.

Position Ergonomically

Position the patient ergonomically to allow for a smooth continuous scan procedure.

Scan Temporaries

If temps are in place scan each temp prosthesis in situ, the established bite and the intaglio surface if modified.

Repeat Scans

Consider repeating scans with the scanner moving in opposite directions. This can help to detect and adjust for scan data for warping artifacts.

Add Soft Tissue

Consider whether an additional soft tissue scan might be required in FP1 cases.